# XTREM Communication Protocol
**Version:** 3.007
**Device:** XTREM / XTREM-S Weighing Module
**Standard:** OIML R76:2006 / EN45501:2015 compliant
---
## Overview
The **XTREM communication protocol** allows a terminal or host controller to exchange data and commands with the XTREM weighing module.
The device transmits and receives ASCII-coded frames over a serial interface (RS232, RS485, Wi-Fi, or Ethernet, depending on configuration).
Since we are using the Ethernet interface we need to provide some kind of DHCP to assign an IP - Address to each XTREM Device.
They then may be addressed via a **broadcast** IP (e.g.: 192.168.4.255/24).
A **Unicast** communication is applied through the device ID inside the XTREM Frame.
---
## Frame Structure
Each message—both query and response—has the same format:
```
STX ID_O ID_D F D_ADDRESS D_L DATA LRC ETX
```
| Field | Description | Length (bytes) |
| :------------ | :------------------------------------------------ | :------------: |
| **STX** | Start of message (ASCII `02h`) | 1 |
| **ID_O** | Sender address (2 ASCII hex chars) | 2 |
| **ID_D** | Destination address (2 ASCII hex chars) | 2 |
| **F** | Function code (`R`, `W`, `E`, etc.) | 1 |
| **D_ADDRESS** | Data address (register, 4 ASCII hex chars) | 4 |
| **D_L** | Data length (2 ASCII hex chars) | 2 |
| **DATA** | Optional data field | Variable |
| **LRC** | Longitudinal Redundancy Check (2 ASCII hex chars) | 2 |
| **ETX** | End of message (ASCII `03h`) | 1 |
In the default configuration the Frame also requires a CRLF at the end.
**Example:**
```
STX ‘1’ ‘7’ ‘0’ ‘1’ ‘R’ ‘0’ ‘1’ ‘0’ ‘1’ ‘0’ ‘0’ LRC ETX CR LF
```
---
## Function Codes
| Function | ASCII | Hex | Description |
| :------- | :---- | :---| :-------------- |
| **R** | R | 52h | Read request |
| **r** | r | 72h | Read response |
| **W** | W | 57h | Write request |
| **w** | w | 77h | Write response |
| **E** | E | 45h | Execute command |
| **e** | e | 65h | Execute response|
### Write response codes
* `'0'` – OK
* `'1'` – Write denied (sealed)
* `'2'` – Read-only register
* `'3'` – Invalid value / out of range
### Execute response codes
* `'0'` – Executed successfully
* `'1'` – Blocked by seal switch
---
## LRC Calculation
The **Longitudinal Redundancy Check** is a one-byte XOR of all message bytes, excluding:
* `STX`
* `ETX`
* The LRC itself
Example C function:
```c
unsigned char LRC(unsigned char* s) {
unsigned char lrc = 0;
while (*s) {
lrc ^= *(s++);
}
return lrc;
}
```
Example Rust function:
```rust
pub fn compute_lrc(data: &[u8]) -> u8 {
data.iter().fold(0u8, |acc, &b| acc ^ b)
}
```
---
## Key Data Registers
| Address (hex) | Description | Type | Notes |
| :------------ | :------------------------- | :--: | :------------------------------------ |
| **0000h** | Serial number | R | Factory-set |
| **0001h** | Device ID | RW | Network address |
| **0007h** | Hardware version | R | Read-only |
| **0008h** | Software version | R | Read-only |
| **0009h** | Sealing switch status | R | `0=UNLOCK`, `1=LOCK` |
| **0010h** | Baud rate | RW | `0–4` → 9600–115200 baud |
| **0013h** | Stream mode output rate | RW | Time in ms |
| **0100h** | Device state | R | Status bits (power, Wi-Fi, ADC, etc.) |
| **0101h** | Weight value | R | Gross weight |
| **0102h** | Tare value | RE | Read/Execute |
| **0103h** | Net weight | R | Gross − Tare |
| **0110h** | ADC counts (instant) | R | Raw ADC data |
| **0111h** | ADC counts (filtered) | R | Post-filter data |
| **1010h** | Stop stream mode | E | Execute |
| **1011h** | Start stream mode (weight) | E | Continuous output |
| **9999h** | Device reset | E | Software reset |
| **EEEEh** | Factory reset | E | Restores defaults |
---
## Post-Factory Reset Configuration
After performing a factory reset, several registers must be configured to ensure proper operation of the device.
Note: The unit may remain set to 'g' after a factory reset, but the values received may be in 'kg' depending on the device configuration.
The following registers should be reviewed and set as required:
1. **Address: 0022 | Max (Max1) (scale maximum capacity)**
The default value for this register may vary between devices (e.g., 60000 or 6000). Set this value to 6000 to ensure correct operation.
2. **Address: 0023 | e (e1) (scale interval)**
3. **Address: 0026 | Decimal position**